Werth Coordinate Measuring Machine with X-Ray Tomography Sensor TomoScope XS FOV Series

Share:


Werth Coordinate Measuring Machine with X-Ray Tomography Sensor TomoScope XS FOV Series

TomoScope XS FOV

Coordinate Measuring Machine with Tomography Sensor
  • Coordinate measuring machine for three-dimensional measurement according to the principle of computed tomography
  • Fast measurement in the field of view of the detector without moving axes with high resolution and high power
  • High availability due to maintenance-free tube in mono-block design
  • Extremely precise air bearing rotary axis for low measurement uncertainty
  • Low space requirement thanks to compact design with integrated control
  • As a result of the low weight, installation is possible almost anywhere
  • Fast amortization through low acquisition costs – X-ray tomography at the price of conventional 3D coordinate measuring machines
  • Standard-compliant calibration for reliable and traceable measurement results, optionally with DAkkS certificate
  • Versatile fields of application, such as plastic workpieces from the packaging industry
  • Software for 3D real-time reconstruction of workpiece geometries during tomography


Option:

  • Automatic determination of the optimal CT setting parameters depending on the respective measuring task
  • OnTheFly tomography for considerable reduction of the measurement time
  • Software package for inspection and material testing tasks
  • Evaluation of STL measuring point clouds regarding voids
  • Automatic separation of the measured data during evaluation
  • Werth Tool Correction (patented) for direct correction of injection molding tools
 
 

Ever since Werth presented the first X-ray tomography machine developed especially for coordinate measuring technology, with optional multisensor systems, the machines in the Werth TomoScope® series have enabled the complete, precise, and non-destructive measurement of workpieces. This technology minimizes first article inspection times for rapid product validation, thus reducing development costs. Special measurement methods also make it possible to monitor processes with high throughput due to short measurement times.
 


Special Features

 

  • CMM technology inside
  • CAA compensation
  • Temperature compensation
  • Pre-calibrated magnifications
  • Fully-automatic measurement
  • Local edge detection using subvoxeling algorithms (patent)
  • No manual influence on the measuring results
  • Integrated artifact correction
  • Accuracy around 2 µm − 20µm (depending on component material and geometry)
 

Principle

X-ray images of the workpiece are taken at various rotated positions. With suitable mathematical methods, they can be used to calculate a complete, high-resolution volumetric model. A patented subvoxeling process determines the measurement points at the material boundaries.

 




 
3D Comparison

The 3D point cloud can be compared directly with the 3D CAD model, imported for example in IGES or STEP format. Each point is automatically associated with the corresponding patch in the CAD model. WinWerth® measuring software generates a color-coded deviation plot for each measurement point in comparison with the specified CAD surface.

 




 
Dimensional Measurements

The points associated with a geometrical feature are selected for analysis, either using automatic segmentation or simply by clicking on the CAD model (patent). WinWerth® calculates the corresponding standard geometrical features (line, cylinder, plane, etc.) and links them into dimensions, generating a measurement report on request. Using virtual planes, measurements can be performed in any cross section of the volume or point cloud (patent pending). Workpieces made of several different materials can be measured using this technique. 

 




 
Machine Design

The design principle of the CT machines, based on proven components from Werth coordinate measuring machines, ensures stability and precision. All TomoScope® and TomoCheck® machines meet the requirements to be considered cabinet X-ray systems.

 


 
Precision in the Sub-Micron Range

The CT measurements can be traced back to the internationally recognized length standard of the German National Metrology Institute (PTB – Physikalisch-Technische Bundesanstalt), using calibrated standards in accordance with VDI 2617 or VDI 2630, with a DAkkS calibration certificate available on request. CT measurements, e.g., on plastics, can be performed with measurement errors of a few micrometers. If precision in the sub-micron range is required for workpieces that are difficult to penetrate radiographically, multisensor systems can help reduce systematic measurement errors due to artifacts, using patented Werth Autocorrection.
 




 
Special Measurement Methods

Special measurement methods, such as OnTheFly CT, Eccentric CT (patent), Multi-ROI CT, Raster Tomography, and Dual-Spectra CT are available to increase the measuring speed or resolution, to expand the measurement range, and to measure workpieces made of multiple materials.

 
 
X-Ray CT Speeds Up First Article Inspection

The measurement time for first article inspection of a workpiece has been reduced from several days, using classical measurement techniques, to a few minutes with CT. Product development can thereby be sped up significantly, which increases cost-effectiveness. With extensive and precise information about the workpiece, mold corrections can be implemented in a fraction of the time.



 


 

WinWerth® Software

The operation of machines with a wide variety of sensors, but also the evaluation of volume data and point clouds are possible with WinWerth® in a unique combination. The Werth image processing software is based on 40 years of experience and is the foundation of probably the most powerful image processing sensors for coordinate measuring machines currently available. Optical distance sensors, conventional styluses in single-point or scanning mode, the Werth Fiber Probe®, X-ray computed tomography or machines with a combination of several sensors are all supported by the uniform concept. Measurement points, 2D images or volume data can also be conveniently evaluated in terms of geometrical characteristics or with part-to-part deviation analysis. PTB-certified evaluation algorithms ensure correct measurement results. All desired information is displayed in the graphic: CAD models with PMI data, voxel volumes, measurement point clouds, colour-coded deviation plots from 3D nominal-actual comparisons, video images, measurement and calculation elements as well as flags with nominal and actual values, tolerances and deviations. In order to meet the most diverse requirements, the software has a modular structure. Various machines can be operated, from simple measuring projectors to complex multi-axis coordinate measuring machines with multi-sensor systems or even X-ray tomography sensors.



 



 
Image Processing Measures almost by Itself

The “intelligence” of the WinWerth® measurement software then takes over, for example, the exact determination of the object area to be captured, the selection of the geometrical element to be measured (e.g. e.g. straight line, circle, corner point) as well as the linking algorithms for determining geometrical characteristics such as distances, angles and diameters.

 


 
Measurement Points Distributed Automatically

Measurement points or scan lines are automatically distributed on the geometry elements to be measured, e.g. as circles, cylinder surface lines, stars or spirals, taking into account the necessary travel paths. In this way, the complete measurement sequence, including evaluation, is first created offline using the CAD model or online with the minimum number of points for the respective geometry element.

 




 
Evaluating Images Perfectly for Optics and Computer Tomography Scan

The evaluation is mainly realised by PC hardware and software. In a first processing step, the image can be improved with image filters (optimising contrast, smoothing surface disturbances). This enables reliable measurements even with difficult edges and rigid scanning in incident light.

 




 
Testing and Changing Made Easy

The feature tree in the WinWerth® user interface also controls the test and change mode, in which programmes can be run step-by-step and changes can be added. A text editor, available in parallel, allows experienced operators to directly enter or change DMIS programme code while teaching in programmes.

 




 
CAD-Online® and CAD-Offline®

Measurement programs can be generated both online and offline using 2D or 3D CAD models. The CAD models are imported in either STEP, native CAD or IGES format. In offline mode a sensor is selected and a patch or combination of several patches is selected on the CAD model. The software computes the necessary actions for the sensor and automatically generates the corresponding segment of the program. The graphic shows the simulated measurement sequence. In online mode the procedure is similar to the offline mode, but the coordinate measuring machine immediately performs each operational step so it can be observed “live.

 




 
Volume Section Sensor

With 2D contour image processing and the associated image processing filters, measurements can also be taken in any cross section of the CT volume or point cloud. Among other things, this makes the measurement of workpieces made of several materials particularly easy.

 




 
Eccentric Sections can be Scanned Tomographically with Multi-ROI CT

With the help of section tomography or ROI tomography (ROI: Region of Interest), parts of the measuring object are measured with high resolution without having to capture the entire measuring object, e.g. , completely with high resolution using Raster Tomography, which is time-consuming and requires a lot of memory. Multi-ROI tomography offers a combination of the benefits of eccentric and sectional tomography scans. Multiple parts with high resolution can also be selected at any position in the measuring object.

 




 
Automatic Burr Detection

A special feature of Werth is the automatic detection and measurement of burrs or chips during the measurement sequence. The result is a colour-coded deviation plot of the burr and the maximum burr length. The deviation display optionally shows only those points where the burr length exceeds the tolerance limits. The burr length along the entire burr can also be displayed numerically via analysis markers. For example, every 0.5 mm a flag is set that contains the maximum local burr length.

 




 
Graphic Displays and Reports

WinWerth® displays all the measured elements along with the selected geometrical characteristics in the 2D or 3D graphics window. The report generator summarizes the various outputs in “Office style.

 


 
Specifications
 
Model TomoScope® XS FOV
General
Machine Type Coordinate measuring machine with CT device
Probing System Compact X-ray sensor
Modes of Operation CNC control
Measuring Software WinWerth®
Operating System MS Windows
Measuring Range
Max. part dimensions for tomography*** For “In the image measurements
from L = 45 mm, Ø = 51 mm
to  L = 120 mm, Ø = 93 mm
Max. distance X-ray source – detector: FDD = 500 mm (19.7″)
Dimensions and Masses [Installation area (without instrument table)]:
Depth 674 mm (26.5″)
Width 1300 mm (51.2″)
Height 1370 mm (53.9″)
Machine Weight 880 kg (1940 Ibs.)
Workpiece Weight mmax 10 kg (22 Ibs.)
– For specified Error MPE 2 kg (4.4 Ibs.)
X-Ray Components
X-Ray Source (according to requirement) Macro focus monoblock tube: 130 kV
 Detector:  
– Surface Area from 60 x 53 mm² to 140 x 110 mm²
– Number of Pixels from (1200 x 1060) pixel to (2800 x 2200) pixel
– Pixel Center Distance 50 µm
Additional Performance Data
Resolution Rotary Axis 0.36″
Supply Data**
Voltage 115 – 230 V AC ±10%
Frequency 50 – 60 Hz
Power Consumption Max. 2000VA
Air Pressure 5.5 – 10 bar
Air Consumption 3000 Nl/h
Permissible Environmental Conditions
Environmental Air Humidity 40% – 70% rel. hum., oil free
Air Contamination Max. 0.05 mg/m³
Operating Temperature 10 – 35 °C (50 – 95 °F)

** Other supply data on request or according to specific countrykit
*** Depending on calibration condition and aspect ratio of the workpieces

 
Maximum Permissible Error MPE (Extract)
For advanced laboratory conditions
CT Sensor 1) P: 5 µm E: (5 + L/75) µm SD: (4 + L/100) µm
No air-conditioning required
CT Sensor 2) P: 7.5 µm E: (7.5 + L/50) µm  
(Where L = measuring length in mm comparable to ISO 10360 and VDI/VDE 2617)


1) ∂ = 20 °C ± 2 K                ∇∂ = 1 K/h                         m ≤ 2 kg (4.4 Ibs.)

2) ∂ = 16°C to 30°C             ∇∂ = 2 K/h, 2 K                m ≤ 2 kg (4.4 lbs.)


Related Video


 Inquiry - Werth Coordinate Measuring Machine with X-Ray Tomography Sensor TomoScope XS FOV Series

GET IN TOUCH
10, Jalan Gemilang 9, Taman Perindustrian Cemerlang, 81800 Ulu Tiram, Johor, Malaysia.
+607-862 6772
STAY CONNECTED